| Model Number |
9735559 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a thermal therapy system being used for a neuro soft tissue ablation procedure.It was reported that intra-operatively, there was an issue with the catheter having a charred tip after the second trajectory.There was artifact during the second trajectory.The surgeon increased irrigation and turned off ablation.The representative replaced the catheter but the laser was used for the remainder of the procedure.The procedure was completed and there was no reported impact to patient outcome.There was a reported delay to the procedure of five minutes due to this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the catheter was returned to the manufacturer for analysis.Analysis found that the reported issue could be confirmed; the returned catheter had a burn mark near the tip.Analysis found that the reported event was related to a mechanical issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|