Product complaint # = > pc-(b)(4).Date sent to the fda: 9/15/2021 h6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.H3 investigation summary: visual analysis of the returned sample determined that one a paper lid and a winding former with an undispensed suture of product code pdp513g were received for evaluation.As per the visual inspection of product received, the suture suffered thermal deformation, including shrinkage, and melting into the winding former; however, in the visual inspection of winding former no damage or defect on the plastic tray could be observed.The slots presented marks and there is evidence which indicates that the needle was placed.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.
|