Product complaint #:(b)(4).Reporter is a j&j representative.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on an unknown date, during inspection of the instruments after going to decontamination process, it was noticed that a trocar was bent and was inadvertently placed in the sharp¿s container, a reaming rod push rod with ball handle was also found bent and has scuff markings on it where it looks like it is getting jammed up in the drills because it is bent, and the whole tip of a depth gauge broke off completely.There was no patient involvement.This complaint involves three (3) devices.This report is for (1) reaming rod push rod with ball handle this report is 2 of 3 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: part number: 03.010.093.Lot number: t932280.Manufacturing site: tuttlingen.Release to warehouse date: 10-dec-2008.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: the reaming rod push rod with ball handle (product code: 03.010.093, lot number: t932280) was received at us customer quality (cq) for investigation.The visual inspection of the received device indicated that the 3.0 mm rod component of the device was bent.Additionally, scratches were observed on the ball handle which were consistent with the field usage.No other issues were identified with the device.Device failure/defect identified? yes.Dimensional inspection: measured dimensions: ø near bent location: conforming.Document/specification review: based on the manufactured date, the following documents reflecting current and manufactured revisions were reviewed as part of this investigation: reaming rod push rod with ball handle.Reaming rod push rod: 3.0mm rod.Based on the review of the above drawings, no design issues contributing to relevant complaint condition were identified.Complaint confirmed? yes.Investigation conclusion: the overall complaint was confirmed for the received device as a component in the device was bent.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|