• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-52
Device Problems Device Displays Incorrect Message (2591); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  malfunction  
Manufacturer Narrative
At this time, the customer has not requested getinge to evaluate the iabp.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that cardiosave intra-aortic balloon pump (iabp) had a gas loss in iab circuit.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) experienced a gas loss in intra-aortic balloon (iab) circuit.There was no patient harm and no adverse event was reported.
 
Manufacturer Narrative
Updated fields: a3, b4, e1 (event site email), g4, g7, h2, h10, h11.Corrected fields: b5, d5, h3, h6 (type of investigation, health effect - impact codes).Testing of actual/suspected device (10): a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed within the iabp log files that the end user had experienced the reported issue on (b)(6) 2021 and continued to use the iabp unit until (b)(6) 2021.The fse tested the iabp unit per factory specifications and was unable to duplicate the reported issue.The iabp unit was then ran on a system trainer for 30 minutes and no alarm was encountered.Subsequently, the fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12269294
MDR Text Key264812063
Report Number2249723-2021-01692
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108438
UDI-Public10607567108438
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
Patient SexPrefer Not To Disclose
-
-