ZIMMER DENTAL IMPL TWIST MP-1 3.75 MM 8 MM; DENTAL IMPLANT Back to Search Results
Model Number 1988
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Pma/510k: k943604.
 
Event or Problem Description
It was reported that the clinician placed implant in the patient's mouth and tried to remove the fixture mount, but it was too hard to remove.The doctor removed the implant and placed another product stocked by the clinic.Tooth site # 13.
 
Additional Manufacturer Narrative
Three impl twist mp-1 3.75 mm 8 mm (1988) and mount with packaging were returned for investigation.Visual evaluation of the as returned product identified signs of use.Pre-existing condition as noted on the per was type ii (moderate) bone density.The implants had been placed on tooth #12, 13 and 15 (universal) and was removed the same day.Functional testing was performed and the two implant with mounts were able to disengage with little effort using normal hand tools but one implant with mount could not be disengaged.X-ray or picture image was not provided.Appropriate documentation was reviewed.Dhr review was completed for the subject lot number 2020060164.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review by lot number (2020060164) was performed for similar events and 1 other similar complaint identified.Based on the available information, device malfunction for one implant/mount has occurred and malfunction for other two implant/mounts has not occurred, therefore, the reported event was confirmed.The following sections have been updated: g6: checked "follow-up." h3: changed "no" to "yes.".
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TWIST MP-1 3.75 MM 8 MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key12270476
Report Number0002023141-2021-02009
Device Sequence Number4440261
Product Code DZE
UDI-Device Identifier00889024013117
UDI-Public(01)00889024013117(17)250610(10)2020060164
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K013494
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 12/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/10/2025
Device Model Number1988
Device Catalogue Number1988
Device Lot Number2020060164
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2021
Supplement Date Received by Manufacturer12/31/2021
Initial Report FDA Received Date08/03/2021
Supplement Report FDA Received Date12/31/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-