Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.The nph low flow valve (909512) was returned for evaluation: device history record (dhr) ¿ the dhr was reviewed and no anomalies that could be associated with the complaint incident were observed.Failure analysis - valve unit was received with distal catheter cut.Fluid and blood were present in the valve.Patency test with air revealed no anomaly.Visual inspection noted a small tear of the valve sole at the inlet connector level, that did not lead to leakage.The valve was pressure/flow tested and found out of specifications: valve underdrains.The valve modulus was removed from its silicone elastomer housing and opened: inspection under magnification showed whitish residues (looking like organic residues) on the diaphragm and around the pin.Root cause -the complaint is verified, the valve as received was out of specifications because of residues impairing the valve mechanism.Each valve is pressure/flow tested at the end of its manufacturing process and valve sn 956864 was tested within specification as shown on the device history records.Valve underdrainage is a known, patient-related complications of valve therapy, as stated in the product instructions for use (ifu).With the available information, no further investigation nor corrective action is deemed required.
|