It was reported that during the arthroscopy procedure, inside the patient, the upper jaw was broken when the disp firstpass str passr self was used.All broken pieces were removed from the patient.The procedure was finished with a smith and nephew backup device.No surgical delay.Patient injuries were not reported.
|
H10 h3, h6 the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.The suture capture is detached and not returned.Product was out of package, no packaging returned.A functional evaluation revealed the needle will deploy when trigger is initiated.Sutures will pass.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.
|