| Model Number |
1988 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Patient weight unknown / not provided.Pma/510k: k943604.
|
| |
|
Event or Problem Description
|
|
It was reported that the doctor placed an implant in a patient's mouth and attempted to remove the fixture mount, but it was too hard to remove.The doctor removed the implant and installed another product that the clinic had in stock.Tooth site # 12.
|
| |
|
Additional Manufacturer Narrative
|
|
One impl twist mp-1 3.75 mm 8 mm (1988) and mount were returned for investigation.Visual evaluation of the as returned product identified signs of use.Pre-existing condition as noted on the per was type iii (low) bone density.The implant had been placed on tooth #12 (universal) and was removed the same day.The reported event could not be recreated when the returned product was functionally tested.The mount was able to disengage with little effort using normal hand tools.X-ray or picture image was not provided.Appropriate documentation was reviewed.Dhr review was completed for the subject lot number 2020060164.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review by lot number (2020060164) was performed for similar events and 1 other similar complaint identified.Based on the available information, device malfunction has not occurred and the reported event was unconfirmed.The following sections have been updated:.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|