MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G247
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hemorrhage/Bleeding (1888); Unspecified Infection (1930)
Event Date 06/22/2021
Event Type  Injury  
Manufacturer Narrative
The product has been received for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection.Additional information indicated that the patient also experienced bleeding.A revision was performed and the crt-d along with the leads were explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
The product has been received for analysis.If information is provided in the future, a supplemental report will be issued.The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection.Additional information indicated that the patient also experienced bleeding.A revision was performed and the crt-d along with the leads were explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key12271979
MDR Text Key264926039
Report Number2124215-2021-23014
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/12/2023
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number258591
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2021
Date Manufacturer Received08/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-