Model Number 7207677
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during surgery, the screw broke down while tightening.The procedure was successfully completed without delay using a back-up device.No other complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A complaint history review concluded this was a repeat issue.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A review of the raw materials found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCR BIORCI-HA 7X30 7MM HEAD STERILE
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key12272244
Report Number1219602-2021-01654
Device Sequence Number17293030
Product Code HWC
UDI-Device Identifier03596010459923
UDI-Public03596010459923
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K002274
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2023
Device Model Number7207677
Device Catalogue Number7207677
Device Lot Number50765111
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/12/2021
Supplement Date Received by Manufacturer09/01/2021
Initial Report FDA Received Date08/04/2021
Supplement Report FDA Received Date09/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-