• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH- AT AXIOM SENSIS HEMO LOW; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH- AT AXIOM SENSIS HEMO LOW; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 6634633
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/17/2021
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the axiom sensis hemo low system.During an interventional procedure, the user reported that the documentation screen went blank and rebooted.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.Assessment does not indicate a system failure or malfunction and no non-conformity was identified.The investigation could not provide a clear result as the system worked again after service intervention.The sensis system is equipped with several sub-components.One of them, namely the dmc (dialog monitor computer) restarted due to a temporary error.The rtc (real time controller) was not affected by this, and the vital signs were visualized to the user continuously.The technician from the local service organization went on site and eliminated the apparent temporary error by unplugging and plugging in the computer boards and installed the software system backup.The system then worked again as specified.Because the system has already reached "end of support" status and no spare parts are in stock nor can further service be ensured, the customer was informed in this regard.It is up to the operator to decide on the possible decommissioning or further use of the system.A possible general fault that would require corrective measures of the installed base could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOM SENSIS HEMO LOW
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key12273321
MDR Text Key264951922
Report Number3004977335-2021-89726
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
PMA/PMN Number
K150493
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6634633
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/19/2021
Initial Date FDA Received08/04/2021
Supplement Dates Manufacturer Received09/30/2021
Supplement Dates FDA Received09/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-