Model Number 2178-63-113
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the stem trial will not fit properly on the 45 mm tibia broach.Stem trial will also not fit properly on the 34 mm and the 31 mm femoral broach.16 mm reamer comes loose from the hudson adapter.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: examination of the returned instrument revealed the o ring was worn and degraded.The investigation did not establish that a broader investigation or corrective action was necessary.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REV TIB 2 DEG BROACH 45MM
Common Device Name
KNEE INSTRUMENT : BROACHES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12273586
Report Number1818910-2021-16943
Device Sequence Number17293161
Product Code HTQ
UDI-Device Identifier10603295096108
UDI-Public10603295096108
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2178-63-113
Device Catalogue Number217863113
Device Lot NumberSO2001727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Initial Date Received by Manufacturer 07/14/2021
Supplement Date Received by Manufacturer09/02/2021
Initial Report FDA Received Date08/04/2021
Supplement Report FDA Received Date09/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-