Catalog Number 8065601187 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
Retinal Detachment (2047)
|
Event Date 07/13/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
No sample has been received at manufacturing for evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.This is second of five reports for this reported event.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A physician reported that when the silicon oil (so) was replaced from perfluoro-n-octane liquid (pfc) in a combination cataract and vitrectomy surgery, the viscosity of silicon oil increased and could not be removed rather, it remained on and migrated under the retina after which the treatment for the rhegmatogenous retinal detachment with the giant tear could not be completed.The surgeon reportedly replaced the perfluoro-n-octane with silicon oil while adding irrigation solution.In that environment, the silicon oil headed over the irrigating solution, causing turbulence in the eye and redetachment of the treated detached retina.The surgeon tried to remove the silicon oil, but it was so viscous that he could not remove it in the back of the eyeball with a backflush needle.He punctured the pars plana with a knife and removed it with a 22g catheter needle, but the oil could not completely removed.An ophthalmic gas, sulfur hexafluoride, was then used and the surgery was completed.The surgery time which was planned for about 90 minutes, was extended to four hours.The patient was transferred to the another hospital for an additional surgery.Additional information has been requested.Additional information received clarified that cataract portions of surgery were performed as usual, a 27g three port was placed and fluid/air exchange was performed.The retina of the giant tear on the superior nasal side was slipped.Perfluoro-n-octane liquid was injected to the bottom of the tear margin during the performed procedure.When silicon oil was injected and the water layer was aspirated with a backflush needle, the perfluoro-n-octane was directly substituted with silicon oil.Once the silicon oil was removed, a mass of silicon oil remained on the retina that did not come upward.It was finally removed by inserting the outer tube of a 22g needle through the perforation wound from a knife and connecting it to viscous fluid control (vfc) aspiration.The sulfur hexafluoride gas was further confirmed to have been used at 20% concentration.Retinal detachment improved with continued prone position.Additional information received further clarified that it was not confirmed if the silicon oil and perfluoro-n-octane liquid remained in the affected eye.The detached retina resolved and post-operative patient condition was good, as a result, a re-operation was not performed.
|
|
Manufacturer Narrative
|
No sample received for this report.Review of the complaint history shows no other similar complaint reported.Review of the manufacturing batch record (mbr) of the lot number provided indicated that product was processed and released according to the product's acceptance criteria.The product is compounded by chemical and thermal extraction, and sterile filtration of oil prior to filling.The product is then filled into its primary packaging components using aseptic processing.The product is terminally sterilized using dry heat.Following sterilization, units are 200% inspected for particulate in the solution and then packaged into pouches and sealed.The product package insert provides preparation and use instructions.The root cause of complaint condition could not be determined.Potential root cause includes: ¿ solution quality issue ¿ highly unlikely, as chemistry and microbiology data is reviewed and verified to meet regulatory requirements prior to release.¿ consumer mishandling - no conclusion can be made regarding the contribution of consumer mishandling as this factor is outside the control of the manufacturing facility.¿ event outside of manufacturer's control (product storage, use, and surgical practice); this could not be confirmed.A comprehensive review was performed including a review of the processes, complaint histories, batch record review, finished product testing results.The review shows that the manufacturing processes were in a state of control.Based on the acceptable mbr review and finished product test results, this lot continues to be acceptable.The manufacturer internal reference number is:(b)(4).
|
|
Search Alerts/Recalls
|
|
|