Model Number N/A
Device Problem Difficult or Delayed Separation (4044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).The inserter was not returned or pictures provided; visual and dimensional evaluations could not be performed.Part and lot identification are necessary for review of device history records, neither were provided.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.  multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2021 - 00433.
 
Event Description
It was reported that the threads stripped on the instrument during the operation.After the surgeon had inserted the shell into acetabulum, he tried to remove the inserter.But couldn't separate the inserter from the shell, so he removed the shell and inserter together with other instruments.Then he separated the inserter from shell successfully in vitro and found the thread stripped and the metal chips had fallen off, doctor use a new shell complete the operation.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN INSERTER
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12274406
MDR Text Key267025423
Report Number0001825034-2021-02305
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age47 YR
-
-