Model Number N/A |
Device Problem
Difficult or Delayed Separation (4044)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).The inserter was not returned or pictures provided; visual and dimensional evaluations could not be performed.Part and lot identification are necessary for review of device history records, neither were provided.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends. multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2021 - 00433.
|
|
Event Description
|
It was reported that the threads stripped on the instrument during the operation.After the surgeon had inserted the shell into acetabulum, he tried to remove the inserter.But couldn't separate the inserter from the shell, so he removed the shell and inserter together with other instruments.Then he separated the inserter from shell successfully in vitro and found the thread stripped and the metal chips had fallen off, doctor use a new shell complete the operation.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|