Model Number 100/597/080CZ
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code Hypoxia (1918)
Date of Event 07/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
Information received a smiths medical tracheostomy|pvc - portex tubes pdt uniperc required intervention to preclude serious impairment.On (b)(6) 2021 0800hrs noticed blue lever (locking mechanism) on trachesotomy was partially lifted and unlocked.Tracheostomy remained in good position, nil issues with ventilation.Vital signs nad.( no actual distress) re-locked.No further issue reported until.Decannulated (b)(6) 2021.
 
Additional Manufacturer Narrative
H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Locking mechanisms were tested (locked and unlocked several times) and no problem was found.Blue locking levers were correctly assembled with adjustable flanges, when closed flanges were firmly attached to tracheostomy tube, when unlocked flanges were smoothly moving.No fault was found with the device regarding the reported customer problem.The cause of the reported problem could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Common Device Name
TRACHEOSTOMY
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
NULL
olomoucka 306
hranice 1, mesto 753 0 1
EZ   753 01
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12276479
Report Number3012307300-2021-08059
Device Sequence Number4495479
Product Code BTO
UDI-Device Identifier15019315107146
UDI-Public15019315107146
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K083031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/597/080CZ
Device Catalogue Number100/597/080CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2021
Supplement Date Received by Manufacturer08/20/2021
Initial Report FDA Received Date08/04/2021
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-