(b)(4).Batch #u5d972.(b)(4).Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pce45a device was returned with no apparent damage and with an ecr45b reload present.The reload was received partially fired 1/10.The manual override door out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, the returned device and reload were tested for functionality in the straight position by resetting and reloading it into the device.The device achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.As part of our quality process, the manufacturing records of this batch was reviewed, and the manufacturing standards were met prior to the release of this batch.The event reported was confirmed and it is related to improper use of the device.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment stop windows resulting in the knife pushing the one piece sled forward and locking the reload.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|