| Model Number |
705004 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation any additional information will be communicated in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that when using a sterilized drill through a sleeve, the sleeve and drill got stuck.The surgeon opened the spare drill tip and used it with another sleeve, but the same event occurred in other hole.After that, the surgery was completed using a 3mm k wire as a substitute and a tap.
|
| |
|
Event or Problem Description
|
|
It was reported that when using a sterilized drill through a sleeve, the sleeve and drill got stuck.The surgeon opened the spare drill tip and used it with another sleeve, but the same event occurred in other hole.After that, the surgery was completed using a 3mm k wire as a substitute and a tap.
|
| |
|
Additional Manufacturer Narrative
|
|
D9 / h3 updated to indicate device was not returned.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence was provided.A device inspection was not possible since the affected device was not returned, and no other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More information as well as the affected device must be available in order to determine the exact root cause.If the device is returned or if any further information is provided, the complaint report will be updated.H3 other text : device disposition is unknown.
|
| |
|