• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH DRILL SLEEVE, LOCKING 3.1MM, SHORT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH DRILL SLEEVE, LOCKING 3.1MM, SHORT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 705004
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of the investigation any additional information will be communicated in a supplemental report.
 
Event or Problem Description
It was reported that when using a sterilized drill through a sleeve, the sleeve and drill got stuck.The surgeon opened the spare drill tip and used it with another sleeve, but the same event occurred in other hole.After that, the surgery was completed using a 3mm k wire as a substitute and a tap.
 
Event or Problem Description
It was reported that when using a sterilized drill through a sleeve, the sleeve and drill got stuck.The surgeon opened the spare drill tip and used it with another sleeve, but the same event occurred in other hole.After that, the surgery was completed using a 3mm k wire as a substitute and a tap.
 
Additional Manufacturer Narrative
D9 / h3 updated to indicate device was not returned.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence was provided.A device inspection was not possible since the affected device was not returned, and no other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More information as well as the affected device must be available in order to determine the exact root cause.If the device is returned or if any further information is provided, the complaint report will be updated.H3 other text : device disposition is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL SLEEVE, LOCKING 3.1MM, SHORT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key12280415
Report Number0008031020-2021-00340
Device Sequence Number19753162
Product Code LXH
UDI-Device Identifier07613327083606
UDI-Public07613327083606
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number705004
Device Catalogue Number705004
Device Lot NumberK048138
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/16/2021
Supplement Date Received by Manufacturer09/30/2021
Initial Report FDA Received Date08/05/2021
Supplement Report FDA Received Date10/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-