|
Analysis found that there was a residue consistent with biological contaminants on the device.Visually, the middle rotating assembly tip was broken at the distal end.The information most likely indicates an interference issue of the diameters during manufacturing, ultimately resulting in the reported event; however, the customer did report catching a pledget, causing the failure prematurely.Functionally, the middle assembly did have significant resistance.If information is provided in the future, a supplemental report will be issued.
|