DEPUY SPINE INC MMG SENSORS 4 PK W GROUND; STIMULATOR, NERVE Back to Search Results
Catalog Number 286202104
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fatigue (1849); Nerve Damage (1979)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional device product code: pdq.Complainant part is not expected to be returned for manufacturer review/investigation.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A device history record (dhr) review was conducted: a review of the receiving inspection (ri) for mmg sensors 4 pk w ground was conducted identifying that lot number 00021565 was released in a single batch.Batch1: lot units were released on 27 oct 2020 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date a patient underwent l4/5 microdiscectomy procedure.Postoperatively the surgeon noted the posterior spine decompressions whereby the patient has developed cauda equina syndrome with bilateral abductor weakness.No further information provided.This report is for one (1) sentio mmg sensors 4 pack with ground.This is report 1 of 5 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: photo investigation: the device was not returned.A photo-investigation was performed on the images provided.Upon inspecting images under notes & attachments section, there were no issues observed with the sentio neuromonitoring which contributes to the adverse event, hence unconfirmed.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Conclusion: the complaint condition can't be confirmed during photo/video investigation.During the investigation, no product design issues or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MMG SENSORS 4 PK W GROUND
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key12281497
Report Number1526439-2021-01611
Device Sequence Number14412925
Product Code ETN
UDI-Device Identifier10705034534749
UDI-Public(01)10705034534749
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number286202104
Device Lot Number21565
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/07/2021
Supplement Date Received by Manufacturer09/08/2021
Initial Report FDA Received Date08/05/2021
Supplement Report FDA Received Date09/27/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MMG BALL TIP PFR; MMG CONTROL BOX; MMG POWER SUPPLY; MMG TABLET
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-