• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID DOMESTIC; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID DOMESTIC; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2021
Event Type  malfunction  
Manufacturer Narrative
The full name of the event site was shortened due to field character limit; the full name is (b)(6) medical center.Testing of actual/suspected device: a getinge field service engineer (fse) called the customer to discuss the issue and found that it was not a service issue.The console was not fully seated in the cart, causing the unit to run in rescue mode.In rescue mode the unit does not charge the batteries.The fse had them seat the console properly into the cart, and the unit proceeded to charge.The fse verified with the customer that the batteries had fully charged overnight.No issue and no onsite service required.A supplemental report will be submitted if this information is provided to us.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) batteries would not charge.It is unknown the circumstances under which the event occurred.However, there was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period (b)(6) 2019 through (b)(6) 2021 was reviewed.There were no triggers identified for the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID DOMESTIC
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12281709
MDR Text Key265195189
Report Number2249723-2021-01728
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-