• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG AS VEGA PS TIBIAL PLATEAU CEMENTED T3; KNEE ENDOPROSTHETICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS VEGA PS TIBIAL PLATEAU CEMENTED T3; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NX055Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report, if available.
 
Event or Problem Description
It was reported that there was an issue with nx055z - as vega ps tibial plateau cemented t3.According to the complaint description, the implant loosened post surgery.A patient's right total knee was being revised for a loose tibial component.Upon exposure it was determined that both the femur & tibial component had debonded from the bone cement on both components.A temporary spacer was inserted until it could be determined that this joint was not infected.The patient is currently scheduled for re-implantation of his right knee on (b)(6) 2021.A revision surgery was necessary.Additional information was not provided.Additional information has been requested but not yet received as of this report.The adverse event is filed under aag reference (b)(4).Involved components: nx031z - as vega ps femoral comp.Cemented f5n r - lot 52360163.
 
Additional Manufacturer Narrative
As of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.Conclusion and measures/preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.For the topic vega loosening a product safety case was initiated.
 
Event or Problem Description
Involved components: nx031z - as vega ps femoral comp.Cemented f5n r - lot 52360163.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS TIBIAL PLATEAU CEMENTED T3
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key12281962
Report Number2916714-2021-00167
Device Sequence Number14638221
Product Code JWH
UDI-Device Identifier04046963810450
UDI-Public4046963810450
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/14/2021
Report Date (Section F) 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNX055Z
Device Catalogue NumberNX055Z
Device Lot Number52346878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/12/2021
Device Age4 YR
Event Location Hospital
Date Report to Manufacturer07/12/2021
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/05/2021
Supplement Report FDA Received Date12/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NX031Z - AS VEGA PS FEMORAL COMP.CEMENTED F5N R; NX031Z - AS VEGA PS FEMORAL COMP.CEMENTED F5N R; NX031Z - AS VEGA PS FEMORAL COMP.CEMENTED F5N R
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Date Report Sent to FDA08/05/2021
-
-