| Model Number |
0220210000 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib tori, onsite.Turning off and on intermitenly.(b)(4).The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause is wear and tear.Poor air flow may affect ventilation of the light source unit.This may result on intermittent loss of light.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|