| Model Number |
MR850 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint mr850 respiratory humidifier is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A distributor in (b)(6) reported, via a fisher and paykel healthcare (f&p) field representative, that the audible alarm of a mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the complaint mr850 respiratory humidifier was returned to the fisher & paykel healthcare (f&p) service center in germany where it was inspected by a trained f&p personnel.The device was performance tested and the audible alarm was checked.Our investigation is based on the information provided by the f&p service technician.Results: performance testing of the complaint mr850 respiratory humidifier revealed that the audible alarm was not functioning.Conclusion: we are unable to determine the cause of the reported event.Our mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 heater base.In addition, the product technical manual states that "all servicing procedures shall be followed by a humidifier functional test and an electrical safety test, to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
|
| |
|
Event or Problem Description
|
|
A distributor in germany reported, via a fisher and paykel healthcare (f&p) field representative, that the audible alarm of a mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
|
| |
|
Search Alerts/Recalls
|