• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ORTHALIGN, INC. ORTHALIGN PLUS SYSTEM; ORTHALIGN PLUS NAVIGATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORTHALIGN, INC. ORTHALIGN PLUS SYSTEM; ORTHALIGN PLUS NAVIGATION UNIT Back to Search Results
Model Number 403001
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This initial report is being filed after the due date as the initial submission was found to have not been received during an attempt to submit the follow-up report.At that time, the error message "error: initial report / prior supplement has not been received.The initial report is missing." alerted orthalign, inc.Of the issue prompting the submission of this initial report.An investigation of the navigation unit and it's event log was performed.The event log was corrupted signifying an electronics issue with the navigation unit.A review of the device history record (dhr) was conducted.The device passed all manufacturing specifications prior to release.Orthalign, inc.Will continue to monitor this issue and take action when/if alert limits are exceeded.
 
Event or Problem Description
An orthalign plus navigation unit set the distal femoral valgus angle so far off that it's been visibly incorrect from a distance (15 degrees or more).Rep checked and assisted nurse with set up and calibration, all steps followed as per technique.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORTHALIGN PLUS SYSTEM
Common Device Name
ORTHALIGN PLUS NAVIGATION UNIT
Manufacturer (Section D)
ORTHALIGN, INC.
120 columbia
suite 500
aliso viejo CA 92656
MDR Report Key12285342
Report Number3007521480-2019-00027
Device Sequence Number14620716
Product Code OLO
UDI-Device Identifier00858704006534
UDI-Public00858704006534
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K162962
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number403001
Device Catalogue Number403001
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/25/2019
Initial Report FDA Received Date08/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-