• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST W/ ECHO 2; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST W/ ECHO 2; SURGICAL MESH Back to Search Results
Catalog Number 5990015
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned for evaluation.The sample evaluation confirms for torn frame material resulting in the detachment of a portion of the frame.No manufacturing anomalies were found.As reported, the issue presented when the user attempted to remove the echo 2 ps through an 8mm trocar.The instructions-for-use (ifu) recommends the use of a 12mm trocar size for the product code used.Evaluation of the sample found that the echo ps 2 frame material was inadvertently torn most likely due to forces applied while removing the frame through a 8mm trocar.Review of manufacturing records confirms product was manufactured to specification, with no indication of a manufacturing related cause for the event reported.To date, this is the only complaint reported for this manufacturing lot of (b)(4) units released for distribution in march, 2021.
 
Event or Problem Description
As reported, during a laparoscopic ventral hernia repair with a 2 cm defect on (b)(6) 2021, the surgeon used a bard/davol ventralight st mesh w/echo 2 positioning system.As reported, a 10mm camera port and x2 8mm ports were used for the mesh insertion/removal.Following placement and fixation of the ventralight st mesh, the surgeon attempted to remove the echo 2 ps frame through the 8 mm port, and one of the "spokes" detached from the frame.The detached portion (spoke) of the frame was completely removed along with the rest of the frame and there was no delay in the case.The surgeon is an experienced user of the device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST W/ ECHO 2
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
MDR Report Key12287314
Report Number1213643-2021-20236
Device Sequence Number3807151
Product Code FTL
UDI-Device Identifier00801741130991
UDI-Public(01)00801741130991
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K170294
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/28/2021
Device Catalogue Number5990015
Device Lot NumberHUFN0601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/19/2021
Initial Report FDA Received Date08/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age41 YR
-
-