• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number SMXP130RH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient underwent an unspecified breast implantation procedure with an artoura breast tissue expander 475cc and experienced deflation on the right side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent removal and replacement with a mentor tissue expander (serial number (b)(4)) on (b)(6) 2021.
 
Additional Manufacturer Narrative
On aug 6 2021, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: that investigation included a review of the manufacturing records for lot-9478420, and a visual evaluation of the device.No leak test was necessary to determine the location of the tear as we observed a tear at the union between the shell and the suture tab in the 6 o¿clock position.When evaluated under a microscope the tear pattern did not reveal parallel striations that are consistent with markings made by a sharp instrument.The results also confirmed that there were no abnormalities with the manufacturing of the tissue expander that would have impacted its performance.Therefore, we were unable to identify the root cause of this tear.While deflation is historically the most common complication related to tissue expanders, the rate of deflation with mentor® tissue expanders remains very low.An incident of this nature may be the result of one or more of the following factors: over-inflation of the tissue expander, use of the tissue expander in emerging surgical techniques not identified in the instructions for use (ifu), continuous and sustained stresses to the tissue expander, vigorous body movement (e.G.Physical exercise) or excessive manipulation/ trauma in the region of the expander.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key12288756
Report Number1645337-2021-08789
Device Sequence Number11911996
Product Code LCJ
UDI-Device Identifier00081317024316
UDI-Public00081317024316
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161176
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/26/2024
Device Model NumberSMXP130RH
Device Catalogue NumberSMXP130RH
Device Lot Number9478420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2021
Initial Date Received by Manufacturer 07/15/2021
Supplement Date Received by Manufacturer08/06/2021
Initial Report FDA Received Date08/06/2021
Supplement Report FDA Received Date08/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-