• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE SYRINGE 1.0ML 29GA 1/2IN; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE SYRINGE 1.0ML 29GA 1/2IN; PISTON SYRINGE Back to Search Results
Catalog Number 326719
Device Problem Material Puncture/Hole (1504)
Patient Problem Needle Stick/Puncture (2462)
Event Date 07/09/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Investigation summary : exec summary - no samples (including photos) were returned therefore bd was not able to duplicate or confirm the customer¿s indicated failure and the root cause is undetermined.A review of the manufacturing records was performed and this is the 1st related complaint for cannula through shield on this lot #.No non-conformances were raised in association with this type of event for this lot concluding all inspections were performed as per the applicable operations and met qc specifications.Capa/sa - based on the above no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) are required at this time.Dhr review - a lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.
 
Event Description
It was reported that 2 bd syringe 1.0ml 29ga 1/2in had shield issues.The following information was provided by the initial reporter : the customer has reported that after using the syringe when disposing them sometimes the needle comes through the side of the plastic cap and also that it can be hard to get the needle up to the top of the orange cap.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 1.0ML 29GA 1/2IN
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12289229
MDR Text Key265588892
Report Number1920898-2021-00870
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K170386
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number326719
Device Lot Number9336456
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-