• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTSTD0620
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-01762, 3005168196-2021-01763, 3005168196-2021-01765, 3005168196-2021-01766.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the anterior communicating artery using penumbra smart coils (smart coils), a penumbra smart coil detachment handle (handle), a non-penumbra microcatheter, and a guidewire.During the procedure, the physician advanced a smart coil into the target vessel using the microcatheter and attempted to detach it using a handle; however, the smart coil failed to detach after several attempts.The physician then manually detached the smart coil; however, the detached smart coil remained in the microcatheter.Therefore, the physician used the pusher assembly of the smart coil to push the smart coil back into the aneurysm.Subsequently, the same issue occurred with three other smart coils.The three smart coils failed to detach using the handle.However, the physician manually detached the three smart coils inside the microcatheter and then used their respective pusher assemblies to push the smart coils back into the aneurysm successfully.The procedure was completed using four additional non-penumbra coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12291334
Report Number3005168196-2021-01764
Device Sequence Number14669693
Product Code HCG
UDI-Device Identifier00814548014678
UDI-Public00814548014678
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/06/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model Number400SMTSTD0620
Device Catalogue Number400SMTSTD0620
Device Lot NumberF98101
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2021
Initial Report FDA Received Date08/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-