• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE SYRINGE 0.5ML 31GA 8MM PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE SYRINGE 0.5ML 31GA 8MM PISTON SYRINGE Back to Search Results
Catalog Number 328509
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/09/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown. The date received by manufacturer has been used for this field. Investigation summary : exec summary - samples were received and an investigation was performed. This is the 1st related complaint for the reported lot number. A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot. Bd was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time. A samples returned - customer returned (1) 0. 5ml insulin syringe in an open polybag from lot# 1011589. Consumer reported found 1 syringe when removed needle shield, needle hub came off with shield. The returned syringe was examined, and it was observed that the needle hub/shield assembly was detached from the barrel. No damage to the barrel tip was observed. Manufacturing ((b)(4)) will be notified of the observed issue. Capa/sa - capa (b)(4) has been opened to address this issue. Dhr review - a lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.
 
Event Description
It was reported that 1 bd syringe 0. 3ml 31g 8mm hub separated from the device. The following information was provided by the initial reporter : the consumer reported that when removed 1 syringe the needle shield hub came off with shield. Date of event : unknown. Sample status : awaiting sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYRINGE 0.5ML 31GA 8MM
Type of DevicePISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12291604
MDR Text Key265569960
Report Number1920898-2021-00873
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/20/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number328509
Device Lot Number1011589
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-