Catalog Number 823162
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cerebral Ventriculomeglia (2133)
Event Date 07/16/2021
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
1 of 3 reports.Other mfg report numbers: 3013886523-2021-00353, 3013886523-2021-00354.A physician reported a hakim valve was implanted to a patient via v-p shunt this year with an unknown setting.Enlargement of ventricles was confirmed, therefore, the set pressure was changed.After several days there was no change in the ventricles and the shunt was replaced on (b)(6) 2021.The patient is in stable condition.The device was used with 823073(serial: unk) and 823072(serial; unk).
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields:  d4, d9, g3, g6, h2, h3, h4, h6, h10 the hakim valve was returned for evaluation.Device history record (dhr) - the product code 823162 with lot 4772341 conformed to the specifications when released to stock.Failure analysis - the valve was visually inspected.Some needle holes were noted through needle chamber.The position of the cam when valve was received was 30mmh2o.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve mechanism, at the time of investigation no functional issues were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Type of Device
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
MDR Report Key12291732
MDR Text Key265565402
Report Number3013886523-2021-00352
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K041296
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number823162
Device Lot Number4772341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2021
Date Manufacturer Received09/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-