| Model Number |
FRED3516-PMA |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Stroke/CVA (1770); Thrombosis/Thrombus (4440)
|
| Date of Event |
05/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and not returned to the manufacturer for evaluation.Post procedural images were not provided.Therefore, the reported event could not be confirmed.The instructions for use (ifu) identifies thrombus and stroke as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a on (b)(6) 2021 a fred stent was successfully implanted to treat a left vertebral artery aneurysm.On (b)(6) 2021, the patient returned to the hospital with stroke like symptoms.The stent had formed thrombus which was removed with an aspiration catheter.The patient was placed on an integrillin drip overnight.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional media was provided by the customer.From the image, there was no measurement of the narrow section of the artery.The image showed an atherosclerotic diseased vessel which can increase the potential for mal apposition.
|
| |
|