• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PROBE SAPHYRE 60 DEG 3MM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PROBE SAPHYRE 60 DEG 3MM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7209685
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Event or Problem Description
It was reported that during a bilateral arthroscopy and bilateral high tibial osteotomy, the electrode of the probe fell into the patient.Although there was no significant delay, it is unknown how the procedure was finished.No other complications were reported.
 
Additional Manufacturer Narrative
A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
H11: corrected data.H6 (health effect - impact code) was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROBE SAPHYRE 60 DEG 3MM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12292235
Report Number3003604053-2021-00240
Device Sequence Number14575361
Product Code GEI
UDI-Device Identifier03596010470287
UDI-Public03596010470287
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K991218
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209685
Device Catalogue Number7209685
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2021
Supplement Date Received by Manufacturer05/17/2023
05/18/2023
Initial Report FDA Received Date08/08/2021
Supplement Report FDA Received Date05/19/2023
05/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-