Catalog Number RF048F
Device Problems Patient-Device Incompatibility (2682); Detachment of Device or Device Component (2907)
Patient Problems Perforation (2001); Foreign Body In Patient (2687)
Event Date 06/02/2021
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event Description
It was reported that during a filter placement procedure, the filter was allegedly found to be fractured.It was further reported that the filter was removed with forceps and the fragment was left behind.The patient status was unknown.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported as unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failures as no objective evidence was provided for review.A definitive root cause for the alleged filter limb detachment and perforation of the inferior vena cava (ivc) could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a filter placement procedure, the filter was allegedly found to be fractured and perforated.It was further reported that the filter was removed with forceps and the fragment was left behind.The patient status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECOVERY FILTER SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12292391
MDR Text Key265542103
Report Number2020394-2021-01436
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K031328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRF048F
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received09/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-