U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. FEMORAL HEAD STERILE PRODUCT; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. FEMORAL HEAD STERILE PRODUCT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: catalog number: 00801803601 lot number: 63186829 brand name: cocr heads.Multiple reports were submitted along with this report: 0001822565-2021-02180.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Device not returned for evaluation.
 
Event or Problem Description
It was reported that the patient was revised due to unknown reasons approximately 5 years post implantation.Additional information on the reported event is unavailable.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.X-ray review by third party hcp states that ap pelvis demonstrates bilateral total hip arthroplasties with symmetric position of the femoral heads and osteopenia.No bony fracture.No radiolucency to suggest loosening of the hardware.Reported event was unable to be confirmed.Review of the device history records identified no related deviations or anomalies during manufacturing.Root cause could not be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD STERILE PRODUCT
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key12293710
Report Number0002648920-2021-00212
Device Sequence Number16939110
Product Code LPH
UDI-Device Identifier00889024144712
UDI-Public(01)00889024144712(17)311025(10)63186829
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K953337
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberN/A
Device Catalogue Number00801803601
Device Lot Number63186829
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/14/2021
Supplement Date Received by Manufacturer10/05/2021
Initial Report FDA Received Date08/09/2021
Supplement Report FDA Received Date10/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight84
-
-