| Model Number |
30263E |
| Medical Device Problem Codes |
Backflow (1064); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: one sample was returned for investigation by the customer.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The sample was primed with saline.A bd secondary set was primed with blue dye and attached to the returned sample.Both sets were allowed to free flow.No backflow was observed.The customer complaint that the tubing had allowed backflow of blood when in use could not be replicated.The root cause could not be determined because the issue could not be replicated.A device history record review for the model lot number was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported 40 inch ext w/2 vlv ports allowed backflow of blood.The following information was provided by the initial reporter: "on friday, july 9, i received a call from anesthsia about iv tubing that had allowed blood to back up and come out of one of the access ports.".
|
| |
|