| Model Number |
1011-0141-04 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a male patient (age unknown), these attached internal handles failed to discharge.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The internal handles were returned to zoll medical corporation; the customer's report was verified.The internal paddles failed to discharge during functional testing due to the shock button being stuck.Due to the age of the internal handles (30+ months) they most likely have met their useful life and exceeded the maximum of autoclave /sterilization cycles.The internal handles were scrapped and replacements were sent to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|