MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9736242 |
Device Problems
Display or Visual Feedback Problem (1184); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 9736226, serial/lot #: version: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system for a functional endoscopic spinal surgery (fess) procedure.It was reported that during a procedure, the site went to change the crosshair option to the dot and the user was exited out of the software.It was noted that they went back and tried to select the patient exam, but the images were missing.The system was rebooted and then became stuck on a black screen.The system was powered down and moved to a different outlet and then rebooted and the issue resolved.There was less than an hour delay to the procedure and no impact on patient outcome.
|
|
Manufacturer Narrative
|
A manufacturer representative went to the site to test the navigation system.The system was performing as intended and no parts were replaced.H6: additional review indicated codes d15, b17, and c20 are no longer applicable to the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: a software investigation analysis was initiated to determine the probable cause of the issue through log analysis.Analysis found that the logs did not capture any core files or segmentation faults on the date of issue.Analysis found that the analysis was inconclusive and probable cause was unable to be determined.H6: fdm b01, fdr c19, and fdc d15 are applicable to the software analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|