Model Number 400SMTHXSFT1H02
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 07/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Potential adverse events in the labeling with the penumbra smart coil system include, but are not limited to dissection, or perforation, aneurysm rupture, intracranial hemorrhage, including death.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-01774.
 
Event or Problem Description
On (b)(6) 2021, the patient underwent a coil embolization procedure to treat a ruptured aneurysm in the left internal carotid artery (ica) and posterior communicating (pcom) artery using penumbra smart coils (smart coil), a non-penumbra microcatheter and a guidewire.During the procedure, the physician implanted eight smart coils in the target location using the microcatheter.While attempting to advance the next smart coil, the physician was unable to advance the smart coil at all past its initial position within its introducer sheath.It was reported that multiple attempts were made to advance the smart coil; however, they were unsuccessful.Therefore, the smart coil was removed.Next, the physician implanted five smart coils in the target location using the microcatheter.After advancing the next smart coil into the target location, the physician attempted to reposition the microcatheter to find a good position for the smart coil.While repositioning the microcatheter, the smart coil and microcatheter extended beyond the aneurysm wall which caused a perforation.The physician then repositioned the microcatheter and smart coil and detached the smart coil in the target location.Additional smart coils were then implanted in the target location to bridge the point of perforation and close the portion of the aneurysm dome.Subsequently, it was noticed that there was a brief and small amount of contrast extravasation which quickly sealed.It was reported there was no change in the patient¿s vitals.The event was considered resolved the same day (b)(6) 2021.The aneurysm perforation was reported to be a serious adverse event with a definite relationship to the smart coil and index procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12297651
Report Number3005168196-2021-01775
Device Sequence Number16939104
Product Code HCG
UDI-Device Identifier00814548015651
UDI-Public00814548015651
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/09/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/25/2025
Device Model Number400SMTHXSFT1H02
Device Catalogue Number400SMTHXSFT1H02
Device Lot NumberF98554
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2021
Initial Report FDA Received Date08/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age87 YR
Date Report Sent to FDA01/01/2005
-
-