• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRECISIONGLIDE NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRECISIONGLIDE NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305111
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd implemented a thorough investigation following our corrective preventative action protocol via capa# (b)(4).This was opened for a trend in reports of ¿needle clogged / blocked¿.The issue is temporary interruption of flow of insulin into the reservoir.Based on the investigation, this issue appears to be related to the use case which involves drawing insulin from a vial into a syringe and then filling the reservoir of the insulin pump via a small port off the body in an iterative process, utilizing existing / ¿off the shelf¿ products rather than a product designed for this use.Bd products are designed to be single use, and because this is an "off the shelf" product, bd specifications do not take into account any requirements specific to the tandem use case.The investigation concludes the complaints are related to the use case and not due to a product nonconformance.
 
Event or Problem Description
It was reported that the bd precisionglide¿ needle was blocked and wouldn't allow insulin to flow through.The following information was provided by the initial reporter: "caller reported that insulin would not go through the needle".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRECISIONGLIDE NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key12297880
Report Number1911916-2021-00799
Device Sequence Number3807512
Product Code FMI
UDI-Device Identifier30382903051114
UDI-Public30382903051114
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K021475
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number305111
Device Catalogue Number305111
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2021
Initial Report FDA Received Date08/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-