• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNPKNA NOVO NORDISK PINK; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNPKNA NOVO NORDISK PINK; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNPKNA
Device Problem Break (1069)
Patient Problems Fatigue (1849); Hyperglycemia (1905); Polydipsia (2604)
Event Date 08/04/2021
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they were experiencing high blood glucose.Blood glucose level at the time of the incident was 490 mg/dl and current blood glucose level was 71 mg/dl.The customer had symptoms like thirst, tired and moderate ketone.Insulin pen had damage on dose dial.The insulin pen will be returned for analysis.
 
Manufacturer Narrative
Serial number: (b)(6).Software version: 3.8.5.Color: pink.Battery life remaining: <10 months.Unit paired successfully to commercial mobile app.The screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No anomalies wit the dose dial noted.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.The injection foot wobbles per design and operating as designed.No physical damage to injection foot or inpen was noted.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNPKNA NOVO NORDISK PINK
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
MDR Report Key12297889
MDR Text Key265772197
Report Number3012822846-2021-00640
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000368
UDI-Public(01)000010862088000368(17)230217
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/17/2023
Device Model NumberMMT-105NNPKNA
Device Catalogue NumberMMT-105NNPKNA
Device Lot NumberB94UB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2021
Date Manufacturer Received09/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age9 YR
-
-