• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SUTURE CUTTER; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SUTURE CUTTER; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72202589
Medical Device Problem Codes Corroded (1131); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: the sample is under evaluation by the manufacturing site.
 
Event or Problem Description
It was reported that, the head position of the suture cutter, does not pass through the wire.It is unknown when the issue was found.Backup device was available to complete the procedure.No delay and no patient injuries were reported.Preliminary results of investigation have concluded that this unit had rust and debris which makes it a reportable event since indicates the device was used in the patient.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual inspection revealed it was returned outside of original packaging.The cutter appears jammed at the distal tip.Rust debris appears at the cutter site.Debris in the cutter blade tip.A functional evaluation was performed on the returned device and found it could not be functioned as intended as cutter jammed at the distal tip.A review of the customer provided image reveals the distal tip of the device appears to be jammed.There was a relationship found between the returned device and the reported incident.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with failure to follow instructions.Factors that could have contributed to the failure include a buildup of corrosion/debris from cleaning or chemical fluid ingress over time.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE CUTTER
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12297904
Report Number1219602-2021-01688
Device Sequence Number14598798
Product Code NBH
UDI-Device Identifier03596010652447
UDI-Public03596010652447
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202589
Device Catalogue Number72202589
Device Lot Number50978
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2021
Supplement Date Received by Manufacturer02/20/2024
Initial Report FDA Received Date08/09/2021
Supplement Report FDA Received Date02/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/22/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight80 KG
-
-