| Model Number |
XIFNT610 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
|
| Date of Event |
06/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant that was placed in site 18 (universal) was removed due to peri-implantitis causing loss of integration.Site was grafted with allograft prior to implant placement.Noted pain at implant site as a result of the event.No further injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay corrected information.The udi number was previously reported incorrectly and has now been updated.The following sections are being reported: d4: unique identifier (udi) number was corrected.H10: narrative/data was updated.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|