• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN LINER +4 NEUT 28IDX58OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN LINER +4 NEUT 28IDX58OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1218-28-458
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 07/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that this patient's hip was revised due to poly wear and instability.The patient's primary 28mm femoral head and pinnacle polyethylene liner were removed and replaced by a 40mm biolox delta ts ceramic revision femoral head and a pinnacle altrx polyethylene liner.Doi: (b)(6) 2003, dor: (b)(6) 2021, left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned acetabular liner confirms the reported poly material wear.There is nothing at all unusual about the wear noted for a device having been implanted for more than 18 years.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN LINER +4 NEUT 28IDX58OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12298138
Report Number1818910-2021-17451
Device Sequence Number3839016
Product Code LPH
UDI-Device Identifier10603295011453
UDI-Public10603295011453
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2003
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-28-458
Device Catalogue Number121828458
Device Lot NumberVX8KY1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2021
Initial Date Received by Manufacturer 07/19/2021
Supplement Date Received by Manufacturer08/09/2021
09/10/2021
Initial Report FDA Received Date08/09/2021
Supplement Report FDA Received Date08/18/2021
09/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
28MM FEMORAL HEAD +5 12/14; PINN LINER +4 NEUT 28IDX58OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-