| Model Number |
470025 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: event 1: it was reported that four sets leaked at the cassette.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Event 1: this report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for further evaluation.As a result of this issue, b.Braun has initiated a voluntary medical device correction for multiple batches of the outlook® pump sets including the batch identified from this complaint.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|