| Model Number |
DTBB1D1 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Failure to Charge (1085); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant products: 694765 lead implanted (b)(6) 2008; 5076-52 lead implanted (b)(6) 2008; 5866-38m adaptor implanted (b)(6) 2015.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) alerted for a charge circuit timeout.It was also noted that the device had an abnormal and significant decrease in battery voltage.The device was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Analysis of the returned device was inconclusive.Hybrid analysis revealed that the device was found to be fully functional with nominal current drain.Charge testing was nominal on battery and external power.S2d analysis determined that the cto occurred during a 30.75 second capacitor formation charge.The cause of this ect was not determined.Eos was not set since the cto did not occur during a therapy charge, but rather during a periodic capacitor formation.No hybrid anomalies identified.High voltage capacitor analysis revealed that no visual or functional anomalies were observed during analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|