COVIDIEN VISI-PRO BALLOON-EXPANDABLE STENT SYSTEM; STENT, ILIAC Back to Search Results
Model Number PXB35-05-17-080
Medical Device Problem Codes Malposition of Device (2616); Activation, Positioning or Separation Problem (2906); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); Unspecified Tissue Injury (4559)
Date of Event 08/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician was attempting to use the visi-pro balloon expandable peripheral stent during procedure to treat the mid left renal artery.The left renal artery had a previous non medtronic stent and imaging showed restenosis.The physician was attempting to extend the stented segment by placing the 5 x 17mm vis-pro stent.A non-medtronic wire and 6fr sheath were used.The sheath was placed in the existing non medtronic stent.The 5x17mm visi-pro stent was loaded on the wire and pushed through the 6fr sheath.The stent became dislodged and came off the balloon.The attempts to push the balloon into the stent were unsuccessful.An attempt was made to snare the stent but while the snare got around the stent the stent deformed a little.It was decided to complete the procedure by leaving the stent and placing the patient on anticoagulants. the un-deployed stent was left distal to the existing stent.The physician made the statement of possible renal infarction of the left kidney but this is unconfirmed.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISI-PRO BALLOON-EXPANDABLE STENT SYSTEM
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key12301215
Report Number2183870-2021-00287
Device Sequence Number3531985
Product Code NIO
UDI-Device Identifier00643169787575
UDI-Public00643169787575
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P030045
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/29/2023
Device Model NumberPXB35-05-17-080
Device Catalogue NumberPXB35-05-17-080
Device Lot NumberB061529
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/02/2021
Supplement Date Received by Manufacturer10/14/2021
Initial Report FDA Received Date08/10/2021
Supplement Report FDA Received Date10/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
-
-