Model Number 1506-40-004
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Fall (1848); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient was revised due to tibia had collapsed on the lateral side and the patient according to the revision surgeon said that the patient must have fallen.Surgeon believed the mcl was compromised.Surgeon removed all components and felt that putting a hinge tka was best for the patient moving forward.Patient had a primary lt tka done 7-1-2019, patient that time had "soft bone" according to the primary surgeon.The surgeon felt that putting in a stem on tibia tray was best for patient and 2 years later the patient needs a revision.Patella was not removed.Doi: (b)(6) 2019 ; dor: (b)(6) 2021 ; left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed, and no evidence of implant fracture, disassociation, or anything indicative of a device nonconformance was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUNE CRS FB TIB BASE SZ 4 CEM
Common Device Name
ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key12301416
Report Number1818910-2021-17506
Device Sequence Number12113245
Product Code JWH
UDI-Device Identifier10603295042785
UDI-Public10603295042785
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K160700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1506-40-004
Device Catalogue Number150640004
Device Lot Number8898837
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/15/2021
Supplement Date Received by Manufacturer09/02/2021
Initial Report FDA Received Date08/10/2021
Supplement Report FDA Received Date09/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CEMENTED STEM 14X30MM; ATTUNE CR FB INSRT SZ 6 7MM; ATTUNE CR FEM LT SZ 6 NAR CEM; ATUNE CRS FB TIB BASE SZ 4 CEM; COMPETITOR CEMENT; ATTUNE CEMENTED STEM 14X30MM; ATTUNE CR FB INSRT SZ 6 7MM; ATTUNE CR FEM LT SZ 6 NAR CEM; ATUNE CRS FB TIB BASE SZ 4 CEM; COMPETITOR CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-