• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC SIGMARP CV TB/IN S3 12.5; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC SIGMARP CV TB/IN S3 12.5; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Model Number 96-2032
Medical Device Problem Codes Fracture (1260); Naturally Worn (2988); Unintended Movement (3026)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/15/2010
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical adverse event received for right knee extensor instability : ligament rupture - subluxation of the patella with disrupted medial retinaculum (slight) event is serious and is considered moderate event is definitely not related to device and is possibly related to procedure.Date of implantation: (b)(6) 2010; date of event (onset): (b)(6) 2010; (right knee).Revision operative notes 05-11-2021 indicate the patient received a right total knee revision due to pain, instability, and feelings of popping and clicking.Upon entering the joint, clear fluid, significant synovitis and pieces of polyethylene were encountered.The tibial poly was determined to be fractured and worn with signs of recurrent subluxation / spin-out.The femoral component was noted to have micromotion and signs of early loosening and thus was revised.The tibial component was also revised without indication of deficiency.The patella was left insitu.The surgery was completed without indication of complication by the surgeon.Treatment: none.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMARP CV TB/IN S3 12.5
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12301880
Report Number1818910-2021-17517
Device Sequence Number5403290
Product Code NJL
UDI-Device Identifier10603295240754
UDI-Public10603295240754
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-2032
Device Catalogue Number962032
Device Lot Number3166771
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/20/2021
Supplement Date Received by Manufacturer09/09/2021
Initial Report FDA Received Date08/10/2021
Supplement Report FDA Received Date09/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MBT CEM KEEL TIB TRAY SZ3; PFC SIGMARP CV TB/IN S3 12.5; PFC*SIGMA C/R NPOR FEM RT SZ 3; PFC*SIGMA/OV/DOME PAT 3PEG,38; MBT CEM KEEL TIB TRAY SZ3; PFC SIGMARP CV TB/IN S3 12.5; PFC*SIGMA C/R NPOR FEM RT SZ 3; PFC*SIGMA/OV/DOME PAT 3PEG,38
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight132
-
-