| Model Number |
IPN035564 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Report number mw5102225.
|
| |
|
Event or Problem Description
|
|
Complaint found in maude database reported as: "a right ij triple lumen catheter was inserted in a pt with severe sepsis.At the conclusion of the procedure, it was noted the guidewire was retained within the catheter and partially in the ivc.The catheter was not free-floating and was successfully retrieved by ir.Unfortunately, guidewires have a known design "flow" in that there was no hard-stop design measures, other than the professional user, to prevent a guidewire from 'slipping away' into the catheter itself.".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Mdr report key (b)(4).Mdr text key (b)(4).Report number mw5102225.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
Complaint found in maude database reported as: "a right ij triple lumen catheter was inserted in a pt with severe sepsis.At the conclusion of the procedure, it was noted the guidewire was retained within the catheter and partially in the ivc.The catheter was not free-floating and was successfully retrieved by ir.Unfortunately, guidewires have a known design "flow" in that there was no hard-stop design measures, other than the professional user, to prevent a guidewire from 'slipping away' into the catheter itself.".
|
| |
|