BECTON DICKINSON TRAY EPID CONT WE18G3.5 C20 LOR5 S/L/L-E; ANESTHESIA CONDUCTION KIT Back to Search Results
Model Number 406047
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Oem manufacture: the manufacturing location for this product is hdq us (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that 2 tray epid cont we18g3.5 c20 lor5 s/l/l-e experienced needle hub holes.The following information was provided by the initial reporter: this week we had two connectors in your epidural kit that had a hole in the side.Ok both instances we were able to break open another type of kit and use theirs, but i wanted to let you know.
 
Event or Problem Description
It was reported that 2 tray epid cont we18g3.5 c20 lor5 s/l/l-e experienced needle hub holes.The following information was provided by the initial reporter: this week we had two connectors in your epidural kit that had a hole in the side.Ok both instances we were able to break open another type of kit and use theirs, but i wanted to let you know.
 
Additional Manufacturer Narrative
H6: investigation summary one photo was provided to our quality team for investigation.Based on the provided photo, the reported failure was not confirmed.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device.As the lot involved in this incident is unknown, a device history review cannot be performed.Based on the available information we are not able to identify a root cause at this time.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAY EPID CONT WE18G3.5 C20 LOR5 S/L/L-E
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12303506
Report Number2243072-2021-02040
Device Sequence Number11754965
Product Code CAZ
UDI-Device Identifier00382904060473
UDI-Public00382904060473
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number406047
Device Catalogue Number406047
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2021
Supplement Date Received by Manufacturer02/02/2022
Initial Report FDA Received Date08/10/2021
Supplement Report FDA Received Date02/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-